Straindiffraktometer EPSILON-MDS
Intracrystalline strain measurements with the strain diffractometer EPSILON-MDS
EPSILON-MDS
© GFZ-Potsdam, Section 2.6- Multi-Detector System with nine Radial-Collimators, Pressure Device and Laser-Extensometer.
Description of the System
|
Distances reactor-sample sample-detector |
100.384m 0.900m |
|
|
Number of collimators (1) |
9 |
|
|
Number of detectors (2) (nine per block) |
81 |
|
|
Goniometer stepping motor-moved (3) hand-moved (4) |
3 x translation 1 x rotation 1 x translation |
|
|
Collimators parameters number of channels angular divergence 2Θ-range collimator length collimator adjustment |
49 20' (1 channel) 82°≤2Θ≤98° 700mm 2 translational, 1 angular |
|
|
Detector parameters length diameter gas pressure |
120mm 10mm 3He 8bar |
|
|
Temperature isolation (10) |
±0.2°C |
|
|
Exstress (6-9) Pressure device, 45° to the incident beam in combination with acoustic emission measurement technique (5) electromechanical force generation bevel gear with backslash-breake |
||
|
Sample cylinders (heatable): 30mm Ø, 60mm length uniaxial load: max. 100kN, 150MPa |
||