Wordmark GFZ Potsdam

Straindiffraktometer EPSILON-MDS

Intracrystalline strain measurements with the strain diffractometer EPSILON-MDS


EPSILON-MDS


© GFZ-Potsdam, Section 2.6
 Multi-Detector System with nine Radial-Collimators, Pressure Device and Laser-Extensometer.


Description of the System
 

Distances

reactor-sample

sample-detector

 

100.384m

0.900m

Number of collimators (1)

9

Number of detectors (2)

(nine per block)

81

Goniometer

stepping

motor-moved (3)

hand-moved (4)

 

3 x translation

1 x rotation

1 x translation

Collimators parameters

number of channels

angular divergence

2Θ-range

collimator length

collimator adjustment

 

49

20' (1 channel)

82°≤2Θ≤98°

700mm

2 translational, 1 angular

Detector parameters

length

diameter

gas

pressure

 

120mm

10mm

3He

8bar

Temperature isolation (10)

±0.2°C

Exstress (6-9)

Pressure device, 45° to the incident beam in combination with acoustic emission measurement technique (5) electromechanical force generation bevel gear with backslash-breake

Sample

cylinders (heatable): 30mm Ø, 60mm length

uniaxial load: max. 100kN, 150MPa

 